產(chǎn)品名稱:Met One Instruments激光粒子計(jì)數(shù)器
產(chǎn)品型號(hào):BT-620
更新時(shí)間:2024-08-17
產(chǎn)品簡介:
美國Met One Instruments BT-620激光粒子計(jì)數(shù)器取樣流量 1cfm (28.3l/min)量程 0-3,000,000個(gè)粒子/立方英尺,Z小測試粒徑 0.3微米,測試粒徑 0.3-2.0微米 (0.1微米遞增設(shè)置) 2.0-10 微米 (0.5微米遞增設(shè)置)
中國電子行業(yè)儀器優(yōu)質(zhì)供應(yīng)商——堅(jiān)融實(shí)業(yè)JETYOO INDUSTRIAL & 堅(jiān)友(上海)測量儀器JETYOO INSTRUMENTS,專業(yè)為中國區(qū)用戶提供的儀器設(shè)備、技術(shù)培訓(xùn)與售后維修服務(wù),Support、銷售Sale、服務(wù)Service,3S公司,為上海華東地區(qū)*一家以技術(shù)為導(dǎo)向的儀器綜合服務(wù)商,是您值得信賴的電子行業(yè)儀器專家。
美國Met One Instruments BT-620 臺(tái)式激光粒子計(jì)數(shù)器BT-620 Bench-Top Particle Counter
產(chǎn)品特點(diǎn): 1cfm流量, 同時(shí)測量和顯示6種粒徑, 實(shí)時(shí)自動(dòng)監(jiān)測流量, 偏差大于+/-10%報(bào)警, 內(nèi)置打印機(jī)
取樣流量 1cfm (28.3l/min)
量程 0-3,000,000個(gè)粒子/立方英尺
zui小測試粒徑 0.3微米
測試粒徑 0.3-2.0微米 (0.1微米遞增設(shè)置)
2.0-10 微米 (0.5微米遞增設(shè)置)
測試及顯示通道 6通道, 在0.3-10微米范圍內(nèi)任選6種粒徑
測試 (取樣) 時(shí)間 1-999秒可調(diào)
取樣間隔時(shí)間 0-999秒可調(diào)
操作按鍵 8個(gè)
讀數(shù)顯示 20位4行LCD
溫/濕度測試 選配G3120探頭可測試溫/濕度測試
儀器內(nèi)存 可保存8000個(gè)測試項(xiàng)
控制軟件 配套軟件可以遠(yuǎn)端控制, 數(shù)據(jù)輸出, 打印/保存數(shù)據(jù)
工作環(huán)境 0-50度
內(nèi)置電池 14.8可充鋰電池, 充電時(shí)間3小時(shí)
續(xù)航時(shí)間 8小時(shí)連續(xù)工作
電源適配器 100-240VAC轉(zhuǎn)16.8VDC通用適配器
儀器外殼 不銹鋼
儀器尺寸 260mm x 203mm x 241mm (L x W x D)
儀器重量 6.3公斤
訂購編號(hào): 描述:
BT-620 包含以下物品
熱敏打印機(jī) (內(nèi)置)
精確取樣頭
清潔過濾器
電源適配器 (充電器)
USB數(shù)據(jù)線
配套軟件
打印紙 (2卷)
選配
G3120 即插即用溫濕度探測頭
80849 ISO 21501-4校正書
美國Met One Instruments BT-620激光粒子計(jì)數(shù)器應(yīng)用領(lǐng)域
Whatever your particular needs, The right choice is only natural. Applications:
Cleanroom Monitoring
Medical Device Mfg.
Filter Testing
Hospitals
Food / Beverage
Aerospace Assembly
Paint Booths
Cosmetic Manufacturing
BT-620 Bench-Top Particle Counter美國Met One Instruments BT-620激光粒子計(jì)數(shù)器性能特點(diǎn)
The BT-610 is a compact battery operated particle counter with six selectable size channels, built-in printer, housed in a portable stainless steel case. Flow rate is 0.1 cfm (2.83 lpm) and is maintained with volumetric flow control for improved accuracy.
Simple Operation
The BT-610 has a large size display with very large characters, audio alarm, and internal printer that work together to verify particulate levels. If the particle count exceeds a pre-set limit, the audible alarm will beep, the screen display will flash, and the printer will print the alarm value. Two programmable count alarm limits may be set.
User Selectable Sizes
Each of the six channels may be individually set by the operator from 0.3 to 10 microns. Custom calibrations are not required.
Data Storage and Connectivity
The BT-610 stores over 8,000 sample events. Sample history events can be viewed or printed and downloaded to a computer using Comet software or simple ASCII terminal commands. Connectivity is built-in for both digital (RS-232, RS-485) and analog output (0-5 VDC). The BT-610 also includes USB interface, and a Memory
Stick port.
Printer
The Printer has two modes of operation.
Print Sample: prints sample data at the end of each sample including alarm, date, time and location.
Recall Data: prints all sample data from first selected date and time to second selected date and time.
Portable Desk Top
The BT-610 provides 6 channels, built-in printer and battery operation in a compact stainless steel case. The built-in battery provides 24 hours of typical (intermittent) operation and 8 hours continuous operation. Connect the AC adapter/charger for longer operation and battery charging.
Temperature / Relative Humidity Option
A plug-and-play option can immediay add ambient temperature and relative humidity capabilities to displayed and logged data .
BT-610 Specifications
PERFORMANCE
Particle Size Range 0.3μm - 10μm, 6 channels
Calibrated Sizes 0.3 μm, 0.5μm, 1.0μm, 2.0μm 5.0μm and 10μm
User Size Settings 0.1μm steps from 0.3μm - 2.0μm
0.5μm steps from 2.0μm - 10μm
Concentration Range 0 – 3,000,000 particles per cubic foot (105,900 particle/L)
Accuracy ± 10% to calibration aerosol
Sensitivity 0.3 μm
Flow Rate 0.1 cfm (2.83 lpm)
Sample Time Adjustable: 1 to 999 seconds
Hold Time Adjustable: 0 to 999 seconds
ELECTRICAL
Light Source Laser Diode, 90mW, 780 nm
Power 14.8V Li-Ion self-contained battery pack – Provides for 8 hours continuous use.
Full recharge approximay 3 hours.
AC Adapter/Charger Li-Ion Charger, 100 – 240 VAC to 16.8 VDC @ 1.8A
Communications USB, RS-232 or RS-485
Standards Meets ISO 21501-4
INTERFACE
Display 20 character x 4 line LCD
Keyboard 8 key membrane type
PHYSICAL
Size Height = 8.75" (22.2 cm) With Handle = 10.25“(26.0 cm)
Width = 6.75" (17.1 cm) Depth = 8.5" (21.5 cm)
Weight 8 lb - 1.6 oz (3.67 kg)
ENVIRONMENTAL
Operating Temperature 0o C to +40o C
Storage Temperature -20o C to +60o C
ACCESSORIES
Supplied Operation Manual (MOI #BT-637S-9800)
USB Cable (MOI #500784)
Comet Software (MOI #80248)
Particle View Software
Battery Charger (MOI #80643)
Iso-kinetic Sample Probe (MOI #G3110)
Zero Particulate Filter (MOI #G3111)
Printer Paper - 2 Rolls (MOI #750514)
Optional RH & Temperature Probe (MOI #G3120)
Flow Meter (MOI #9801)
Serial Cable (MOI #550065)
ISO 21501-4 Calibration (MOI #80849)